_____________________
General Information About GCMS
Gas chromatography-mass spectrometry
(GC-MS) is a method that combines the features of gas-liquid
chromatography and mass spectrometry to identify different substances within a
test sample. Applications of GC-MS include drug detection, fire
investigation, environmental analysis, explosives investigation, and
identification of unknown samples. GC/MS can also be used in airport security to
detect substances in luggage or on human beings. Additionally, it can identify
trace elements in materials that were previously thought to have disintegrated
beyond identification.
The GC-MS has been widely heralded as a
"gold standard" for forensic substance identification because it is used to
perform a specific test. A specific test positively identifies the actual
presence of a particular substance in a given sample. A non-specific test
merely indicates that a substance falls into a category of substances. Although
a non-specific test could statistically suggest the identity of the substance,
this could lead to false positive identification.

History
of GCMS
The use of a mass spectrometer as the
detector in gas chromatography was developed during the 1950s by Roland Gohlke
and Fred McLafferty.[1][2] These sensitive devices
were bulky, fragile, and originally limited to laboratory settings. The
development of affordable and miniaturized computers has helped in the
simplification of the use of this instrument, as well as allowed great
improvements in the amount of time it takes to analyze a sample. In 1996 the
top-of-the-line high-speed GC-MS units completed analysis of fire accelerants in
less than 90 seconds, whereas first-generation GC/MS would have required at
least 16 minutes. This has led to their widespread adoption in a number of
fields.
Instrumentation
The insides of the
GC-MS, with the column of the gas chromatograph in the oven on the right.
The GC-MS is composed of two major
building blocks: the gas chromatograph and the mass spectrometer. The gas
chromatograph utilizes a capillary column which depends on the column's
dimensions (length, diameter, film thickness) as well as the phase properties
(e.g. 5% phenyl polysiloxane). The difference in the chemical properties between
different molecules in a mixture will separate the molecules as the sample
travels the length of the column. The molecules take different amounts of time
(called the retention time) to come out of (elute from) the gas chromatograph,
and this allows the mass spectrometer downstream to capture, ionize, accelerate,
deflect, and detect the ionized molecules separately. The mass spectrometer does
this by breaking each molecule into ionized fragments and detecting these
fragments using their mass to charge ratio.

GC-MS schematic
These two components, used together, allow
a much finer degree of substance identification than either unit used
separately. It is not possible to make an accurate identification of a
particular molecule by gas chromatography or mass spectrometry alone. The mass
spectrometry process normally requires a very pure sample while gas
chromatography using a traditional detector (e.g. Flame Ionization Detector)
detects multiple molecules that happen to take the same amount of time to travel
through the column (i.e. have the same retention time) which results in
two or more molecules to co-elute. Sometimes two different molecules can also
have a similar pattern of ionized fragments in a mass spectrometer (mass
spectrum). Combining the two processes makes it extremely unlikely that two
different molecules will behave in the same way in both a gas chromatograph and
a mass spectrometer. Therefore when an identifying mass spectrum appears at a
characteristic retention time in a GC-MS analysis, it typically lends to
increased certainty that the analyte of interest is in the sample.
Split/Splitless GC-MS inlets
Samples are introduced to the column via
an inlet. This inlet is typically injection through a septum. Once in the inlet,
the heated chamber acts to volatilize the sample. In a split system, a constant
flow of carrier gas moves through the inlet. A portion of the carrier gas flow
acts to transport the sample into the column. Another portion of the carrier gas
flow gets directed to purge the inlet of any sample following injection (septum
purge). Yet another portion of the flow is directed through the split vent in a
set ratio known as the split ratio. In a splitless system, the advantage is that
a larger amount of sample is introduced to the column. However, a split system
is preferred when the detector is sensitive to trace amounts of analyte and
there is concern about overloading the column.
Purge and
Trap GC-MS
For the analysis of volatile compounds a
Purge and Trap (P&T) concentrator system may be used to introduce samples. The
target analytes are extracted and mixed with water and introduced into an
airtight chamber. An inert gas such as Nitrogen (N2) is bubbled
through the water; this is known as purging. The volatile compounds move into
the headspace above the water and are drawn along a pressure gradient (caused by
the introduction of the purge gas) out of the chamber. The volatile compounds
are drawn along a heated line onto a 'trap'. the trap is a column of adsorbent
material at ambient temperature that holds the compounds by returning them to
the liquid phase. The trap is then heated and the sample compounds are
introduced to the GC-MS column via a volatiles interface, which is a split inlet
system. P&T GC-MS is particularly suited to volatile organic compounds (VOCs)
and BTEX compounds (aromatic compounds associated with petroleum).[3]
Types of
Mass Spectrometer Detectors
The most common type of mass spectrometer
(MS) associated with a gas chromatograph (GC) is the quadrupole mass
spectrometer, sometimes referred to by the Hewlett-Packard (now Agilent) trade
name "Mass Selective Detector" (MSD). Another relatively common detector is the
ion trap mass spectrometer. Additionally one may find a magnetic sector mass
spectrometer, however these particular instruments are expensive and bulky and
not typically found in high-throughput service laboratories. Other detectors may
be encountered such as time of flight (TOF), tandem quadrupoles (MS-MS) (see
below), or in the case of an ion trap MSn where n indicates the
number mass spectrometry stages.
GCMS Analysis
A mass spectrometer is typically utilized
in one of two ways: Full Scan or Selective Ion Monitoring (SIM). The typical
GC/MS instrument is capable of performing both functions either individually or
concomitantly, depending on the setup of the particular instrument.
Full scan MS
When collecting data in the full scan
mode, a target range of mass fragments is determined and put into the
instrument's method. An example of a typical broad range of mass fragments to
monitor would be m/z 50 to m/z 400. The determination of what
range to use is largely dictated by what one anticipates being in the sample
while being cognizant of the solvent and other possible interferences. A MS
should not be set to look for mass fragments too low or else one may detect air
(found as m/z 28 due to nitrogen), carbon dioxide (m/z 44) or
other possible interferences. Additionally if one is to use a large scan range
then sensitivity of the instrument is decreased due to performing fewer scans
per second since each scan will have to detect a wide range of mass fragments.
Full scan is useful in determining unknown
compounds in a sample. It provides more information than SIM when it comes to
confirming or resolving compounds in a sample. During instrument method
development it may be common to first analyze test solutions in full scan mode
to determine the retention time and the mass fragment fingerprint before moving
to a SIM instrument method.
Selected ion
monitoring
In selected ion monitoring (SIM) certain
ion fragments are entered into the instrument method and only those mass
fragments are detected by the mass spectrometer. The advantages of SIM are that
the detection limit is lower since the instrument is only looking at a small
number of fragments (e.g. three fragments) during each scan. More scans can take
place each second. Since only a few mass fragments of interest are being
monitored, matrix interferences are typically lower. To additionally confirm the
likelihood of a potentially positive result, it is relatively important to be
sure that the ion ratios of the various mass fragments are comparable to a known
reference standard.
Types of
Ionization
After the molecules travel the length of
the column, pass through the transfer line and enter into the mass spectrometer
they are ionized by various methods with typically only one method being used at
any given time. Once the sample is fragmented it will then be detected, usually
by an electron multiplier diode, which essentially turns the ionized mass
fragment into an electrical signal that is then detected.
The ionization technique chosen is
independent of using Full Scan or SIM.
Electron
Ionization
By far the most common and perhaps
standard form of ionization is electron ionization (EI). The molecules enter
into the MS (the source is a quadrupole or the ion trap itself in an ion trap
MS) where they are bombarded with free electrons emitted from a filament, not
much unlike the filament one would find in a standard light bulb. The electrons
bombard the molecules causing a hard ionization that fragments the molecule, and
the way in which a molecule fragment is usually typical for all EI techniques.
Chemical
Ionization
In chemical ionization a reagent gas,
typically methane or ammonia is introduced into the mass spectrometer. Depending
on the technique (positive CI or negative CI) chosen, this reagent gas will
interact with the electrons and analyte and cause a 'soft' ionization of the
molecule of interest. A softer ionization fragments the molecule to a lower
degree than the hard ionization of EI. One of the main benefits of using
chemical ionization is that a mass fragment closely corresponding to the
molecular weight of the analyte of interest is produced.
Positive Chemical
Ionization
In Positive Chemical Ionization (PCI) the
reagent gas interacts with the target molecule, most often with a proton
exchange. This produces the species in relatively high amounts.
Negative Chemical
Ionization
In Negative Chemical Ionization (NCI) the
reagent gas decreases the impact of the free electrons on the target analyte.
This decreased energy typically leaves the fragment in great supply.
The primary goal of instrument analysis is
to quantify an amount of substance. This is done by comparing the relative
concentrations among the atomic masses in the generated spectrum. Two kinds of
analysis are possible, comparative and original. Comparative analysis
essentially compares the given spectrum to a spectrum library to see if its
characteristics are present for some sample in the library. This is best
performed by a computer because there are a myriad of visual distortions that
can take place due to variations in scale. Computers can also simultaneously
correlate more data (such as the retention times identified by GC), to more
accurately relate certain data.
Another method of analysis measures the
peaks in relation to one another. In this method, the tallest peak is assigned
100% of the value, and the other peaks being assigned proportionate values. All
values above 3% are assigned. The total mass of the unknown compound is normally
indicated by the parent peak. The value of this parent peak can be used to fit
with a chemical formula containing the various elements which are believed to be
in the compound. The isotope pattern in the spectrum, which is unique for
elements that have many isotopes, can also be used to identify the various
elements present. Once a chemical formula has been matched to the spectrum, the
molecular structure and bonding can be identified, and must be consistent with
the characteristics recorded by GC/MS. Typically, this identification done
automatically by programs which come with the instrument, given a list of the
elements which could be present in the sample.
A “full spectrum” analysis considers all
the “peaks” within a spectrum. Conversely, selective ion monitoring (SIM) only
monitors selected peaks associated with a specific substance. This is done on
the assumption that at a given retention time, a set of ions is characteristic
of a certain compound. This is a fast and efficient analysis, especially if the
analyst has previous information about a sample or is only looking for a few
specific substances. When the amount of information collected about the ions in
a given gas chromatographic peak decreases, the sensitivity of the analysis
increases. So, SIM analysis allows for a smaller quantity of a compound to be
detected and measured, but the degree of certainty about the identity of that
compound is reduced.
GC-MS/MS
When a second phase of mass fragmentation
is added, for example using a second quadrupole in a quadrupole instrument, it
is called MS/MS or Tandem MS. Tandem mass spectrometry (MS/MS) is a more
powerful technique to quantitate low levels of target compounds in the presence
of a high sample matrix background.
The first quadrupole (Q1) is connected
with a collision cell (q2) and another quadrupole (Q3). Both quadrupoles can be
used in scanning or static mode, depending on the type of MS/MS analysis being
performed. Types of analysis include product ion scan, precursor ion scan,
Single Reaction Monitoring (SRM) and Multiple Reaction Monitoring (MRM) and
Neutral Loss Scan. For example: When Q1 is in static mode (looking at one mass
only as in SIM), and Q3 is in scanning mode, one obtains a so-called product ion
spectrum (also called "daughter spectrum"). From this spectrum, one can select a
prominent product ion which can be the product ion for the chosen precursor ion.
The pair is called a "transition" and forms the basis for SRM (MRM is sometimes
used as term). SRM is highly specific and virtually eliminates matrix
background.
Applications
for the GCMS
Environmental Monitoring and Cleanup
GC-MS is becoming the tool of choice for
tracking organic pollutants in the environment. The cost of GC-MS equipment has
decreased significantly, and the reliability has increased at the same time,
which has contributed to its increased adoption in environmental studies. There
are some compounds for which GC-MS is not sufficiently sensitive, including
certain pesticides and herbicides, but for most organic analysis of
environmental samples, including many major classes of pesticides, it is very
sensitive and effective.
Criminal
Forensics
GC-MS can analyze the particles from a
human body in order to help link a criminal to a crime. The analysis of fire
debris using GC-MS is well established, and there is even an established
American Society for Testing Materials (ASTM) standard for fire debris analysis.
GCMS/MS is especially useful here as samples often contain very complex matrices
and results, used in court, need to be highly accurate.
Law
Enforcement
GC-MS is increasingly used for detection
of illegal narcotics, and may eventually supplant drug-sniffing dogs.[1]
It is also commonly used in forensic toxicology to find drugs and/or poisons in
biological specimens of suspects, victims, or the deceased.
Security
A post-September
11 development, explosive detection systems have become a part of all US
airports. These systems run on a host of technologies, many of them based on
GC-MS. There are only three manufacturers certified by the FAA to provide these
systems, one of which is Thermo Detection (formerly Thermedics), which produces
the EGIS, a GC-MS-based line of explosives detectors. The other two
manufacturers are Barringer Technologies, now owned by Smith's Detection Systems
and Ion Track Instruments, part of General Electric Infrastructure Security
Systems.
Food, Beverage and Perfume
Analysis
Foods and beverages contain numerous
aromatic compounds, some naturally present in the raw materials and some forming
during processing. GC-MS is extensively used for the analysis of these compounds
which include esters, fatty acids, alcohols, aldehydes, terpenes etc. It is also
used to detect and measure contaminants from spoilage or adulteration which may
be harmful and which is often controlled by governmental agencies, for example
pesticides.
Astrochemistry
Several GC-MS have left earth. Two were
brought to Mars by the Viking program.[4] Venera 11
and 12 and Pioneer Venus analysed the atmosphere of Venus with GC-MS.[5]
The Huygens probe of the Cassini-Huygens mission landed one GC-MS on
Saturn's largest moon, Titan.[6] The material in the
comet 67P/Churyumov-Gerasimenko will be analysed by the Rosetta mission with a
chiral GC-MS in 2014.[7]
Medicine
In combination with isotopic labeling of
metabolic compounds, the GC-MS is used for determining metabolic activity. Most
applications are based on the use of 13C as the labeling and the
measurement of 13C/12C ratios with an isotope ratio
mass spectrometer (IRMS); an MS with a detector designed to measure a
few select ions and return values as ratios.
See also
References
-
Gohlke, R. S. (1959).
"Time-of-Flight Mass Spectrometry and Gas-Liquid Partition Chromatography".
Analytical Chemistry 31: 535.
-
Gohlke, R (1993). "Early gas
chromatography/mass spectrometry". Journal of the American Society for Mass
Spectrometry 4: 367.
-
"Optimizing the Analysis of Volatile
Organic Compounds - Technical Guide" Restek Corporation, Lit. Cat. 59887A
-
The Development of the Viking GCMS
-
V. A. Krasnopolsky, V. A. Parshev
(1981). "Chemical composition of the atmosphere of Venus". Nature
292: 610–613.
-
H. B. Niemann, S. K. Atreya, S. J.
Bauer, G. R. Carignan, J. E. Demick, R. L. Frost, D. Gautier, J. A. Haberman,
D. N. Harpold, D. M. Hunten, G. Israel, J. I. Lunine, W. T. Kasprzak, T. C.
Owen, M. Paulkovich, F. Raulin, E. Raaen, S. H. Way (2005). "The abundances of
constituents of Titan’s atmosphere from the GCMS instrument on the Huygens
probe". Nature 438: 77–9–784.
-
Goesmann F, Rosenbauer H, Roll R,
Bohnhardt H (2005). "COSAC onboard Rosetta: A bioastronomy experiment for the
short-period comet 67P/Churyumov-Gerasimenko". Astrobiology 5 (5): 622–631.